Jump to content

Artificial heart valve

From Wikipedia, the free encyclopedia
(Redirected from Bioprosthetic heart valve)
Artificial heart valve
Different types of artificial heart valves[1]
SpecialtyCardiology

An artificial heart valve is a one-way valve implanted into a person's heart to replace a heart valve that is not functioning properly (valvular heart disease). Artificial heart valves can be separated into three broad classes: mechanical heart valves, bioprosthetic tissue valves and engineered tissue valves.

The human heart contains four valves: tricuspid valve, pulmonary valve, mitral valve and aortic valve. Their main purpose is to keep blood flowing in the proper direction through the heart, and from the heart into the major blood vessels connected to it (the pulmonary artery and the aorta). Heart valves can malfunction for a variety of reasons, which can impede the flow of blood through the valve (stenosis) and/or let blood flow backwards through the valve (regurgitation). Both processes put strain on the heart and may lead to serious problems, including heart failure. While some dysfunctional valves can be treated with drugs or repaired, others need to be replaced with an artificial valve.[2]

Background

[edit]
3D Medical Animation still shot of Artificial Heart Valve
3D Medical Animation still shot of Artificial Heart Valve

A heart contains four valves (tricuspid, pulmonary, mitral and aortic valves) which open and close as blood passes through the heart.[3] Blood enters the heart in the right atrium and passes through the tricuspid valve to the right ventricle. From there, blood is pumped through the pulmonary valve to enter the lungs. After being oxygenated, blood passes to the left atrium, where is it pumped through the mitral valve to the left ventricle. The left ventricle pumps blood to the aorta through the aortic valve.

There are many potential causes of heart valve damage, such as birth defects, age related changes, and effects from other disorders, such as rheumatic fever and infections causing endocarditis. High blood pressure and heart failure which can enlarge the heart and arteries, and scar tissue can form after a heart attack or injury.[4]

The three main types of artificial heart valves are mechanical, biological (bioprosthetic/tissue), and tissue-engineered valves. In the US, UK and the European Union, the most common type of artificial heart valve is the bioprosthetic valve. Mechanical valves are more commonly used in Asia and Latin America.[citation needed] Companies that manufacture heart valves include Edwards Lifesciences,[5] Medtronic,[6] Abbott (St. Jude Medical),[7] CryoLife,[8] and LifeNet Health.[9]

Mechanical valves

[edit]

Mechanical valves come in three main types – caged ball, tilting-disc and bileaflet – with various modifications on these designs.[10] Caged ball valves are no longer implanted.[11] Bileaflet valves are the most common type of mechanical valve implanted in patients today.[12]

Caged ball valves

[edit]
Caged ball valve

The first artificial heart valve was the caged ball valve, a type of ball check valve, in which a ball is housed inside a cage. When the heart contracts and the blood pressure in the chamber of the heart exceeds the pressure on the outside of the chamber, the ball is pushed against the cage and allows blood to flow. When the heart finishes contracting, the pressure inside the chamber drops and the ball moves back against the base of the valve forming a seal.

In 1952, Charles A. Hufnagel implanted caged ball heart valves into ten patients (six of whom survived the operation), marking the first success in prosthetic heart valves.[citation needed] A similar valve was invented by Miles 'Lowell' Edwards and Albert Starr in 1960, commonly referred to as the Starr-Edwards silastic ball valve.[13] This consisted of a silicone ball enclosed in a methyl metacrylate cage welded to a ring. The Starr-Edwards valve was first implanted in a human on August 25, 1960, and was discontinued by Edwards Lifesciences in 2007.[13]

Caged ball valves are strongly associated with blood clot formation, so people who have one required a high degree of anticoagulation, usually with a target INR of 3.0–4.5.[14]

Tilting-disc valves

[edit]
tilting-disc valve

Introduced in 1969, the first clinically available tilting-disc valve was the Bjork-Shiley valve.[15] Tilting‑disc valves, a type of swing check valve, are made of a metal ring covered by an ePTFE fabric. The metal ring holds, by means of two metal supports, a disc that opens when the heart beats to let blood flow through, then closes again to prevent blood flowing backwards. The disc is usually made of an extremely hard carbon material (pyrolytic carbon), enabling the valve to function for years without wearing out.[citation needed]

Bileaflet valves

[edit]
Bileaflet valve

Introduced in 1979, bileaflet valves are made of two semicircular leaflets that revolve around struts attached to the valve housing. With a larger opening than caged ball or tilting-disc valves, they carry a lower risk of blood clots. They are, however, vulnerable to blood backflow.[citation needed]

Advantages of mechanical valves

[edit]

The major advantage of mechanical valves over bioprosthetic valves is their greater durability.[16] Made from metal and/or pyrolytic carbon,[10] they can last 20–30 years.[16]

Disadvantages of mechanical valves

[edit]

One of the major drawbacks of mechanical heart valves is that they are associated with an increased risk of blood clots. Clots formed by red blood cell and platelet damage can block blood vessels leading to stroke. People with mechanical valves need to take anticoagulants (blood thinners), such as warfarin, for the rest of their life.[16] Mechanical heart valves can also cause mechanical hemolytic anemia, a condition where the red blood cells are damaged as they pass through the valve.[citation needed] Cavitation, the rapid formation of microbubbles in a fluid such as blood due to a localized drop of pressure, can lead to mechanical heart valve failure,[17] so cavitation testing is an essential part of the valve design verification process.

Many of the complications associated with mechanical heart valves can be explained through fluid mechanics. For example, blood clot formation is a side effect of high shear stresses created by the design of the valves. From an engineering perspective, an ideal heart valve would produce minimal pressure drops, have small regurgitation volumes, minimize turbulence, reduce prevalence of high stresses, and not create flow separations in the vicinity of the valve.[citation needed]

Implanted mechanical valves can cause foreign body rejection. The blood may coagulate and eventually result in a hemostasis. The usage of anticoagulation drugs will be interminable to prevent thrombosis.[18][non-primary source needed]

Bioprosthetic tissue valves

[edit]

Bioprosthetic valves are usually made from animal tissue (heterograft/xenograft) attached to a metal or polymer support.[11] Bovine (cow) tissue is most commonly used, but some are made from porcine (pig) tissue.[19][non-primary source needed] The tissue is treated to prevent rejection and calcification.[citation needed]

Alternatives to animal tissue valves are sometimes used, where valves are used from human donors, as in aortic homografts and pulmonary autografts. An aortic homograft is an aortic valve from a human donor, retrieved either after their death or from a heart that is removed to be replaced during a heart transplant.[12] A pulmonary autograft, also known as the Ross procedure, is where the aortic valve is removed and replaced with the patient's own pulmonary valve (the valve between the right ventricle and the pulmonary artery). A pulmonary homograft (a pulmonary valve taken from a cadaver) is then used to replace the patient's own pulmonary valve. This procedure was first performed in 1967 and is used primarily in children, as it allows the patient's own pulmonary valve (now in the aortic position) to grow with the child.[12]

Advantages of bioprosthetic heart valves

[edit]

Bioprosthetic valves are less likely than mechanical valves to cause blood clots, so do not require lifelong anticoagulation. As a result, people with bioprosthetic valves have a lower risk of bleeding than those with mechanical valves.[16]

Disadvantages of bioprosthetic heart valves

[edit]

Tissue valves are less durable than mechanical valves, typically lasting 10–20 years.[20] This means that people with bioprosthetic valves have a higher incidence of requiring another aortic valve replacement in their lifetime.[16] Bioprosthetic valves tend to deteriorate more quickly in younger patients.[21]

In recent years, scientists have developed a new tissue preservation technology, with the aim of improving the durability of bioprosthetic valves. In sheep and rabbit studies, tissue preserved using this new technology had less calcification than control tissue.[22] A valve containing this tissue is now marketed, but long-term durability data in patients are not yet available.[23][non-primary source needed]

Current bioprosthetic valves lack longevity, and will calcify over time.[24] When a valve calcifies, the valve cusps become stiff and thick and cannot close completely.[24] Moreover, bioprosthetic valves can't grow with or adapt to the patient: if a child has bioprosthetic valves they will need to get the valves replaced several times to fit their physical growth.[24]

Tissue-engineered valves

[edit]

For over 30 years researchers have been trying to grow heart valves in vitro.[25] These tissue‑engineered valves involve seeding human cells on to a scaffold.[25] The two main types of scaffold are natural scaffolds, such as decellularized tissue, or scaffolds made from degradable polymers.[26] The scaffold acts as an extracellular matrix, guiding tissue growth into the correct 3D structure of the heart valve.[26][25] Some tissue-engineered heart valves have been tested in clinical trials,[26] but none are commercially available.

Tissue engineered heart valves can be person-specific and 3D modeled to fit an individual recipient[27] 3D printing is used because of its high accuracy and precision of dealing with different biomaterials.[27] Cells that are used for tissue engineered heart valves are expected to secrete the extracellular matrix (ECM).[24] Extracellular matrix provides support to maintain the shape of the valves and determines the cell activities.[28]

Scientists can follow the structure of heart valves to produce something that looks similar to them, but since tissue engineered valves lack the natural cellular basis, they either fail to perform their functions like natural heart valves, or function when they are implanted but gradually degrade over time.[citation needed] An ideal tissue engineered heart valve would be non-thrombogenic, biocompatible, durable, resistant to calcification, grow with the surrounding heart, and exhibit a physiological hemodynamic profile.[29] To achieve these goals, the scaffold should be carefully chosen—there are three main candidates: decellularized ECM (xenografts or homografts), natural polymers, and synthetic polymers.[29]

Differences between mechanical and tissue valves

[edit]

Mechanical and tissue valves are made of different materials. Mechanical valves are generally made of titanium and carbon.[30] Tissue valves are made up of human or animal tissue. The valves composed of human tissue, known as allografts or homografts, are from donors' human hearts.[30]

Mechanical valves can be a better choice for younger people and people at risk of valve deterioration due to its durability. It is also preferable for people who are already taking blood thinners and people who would be unlikely to tolerate another valve replacement operation.[citation needed]

Tissue valves are better for older age groups as another valve replacement operation may not be needed in their lifetime. Due to the risk of forming blood clots for mechanical valves and severe bleeding as a major side effect of taking blood-thinning medications, people who have a risk of blood bleeding and are not willing to take warfarin may also consider tissue valves. Other patients who may be more suitable for tissue valves are people who have other planned surgeries and unable to take blood-thinning medications. People who plan to become pregnant may also consider tissue valves as warfarin causes risks in pregnancy.[citation needed]

Functional requirements of artificial heart valves

[edit]

An artificial heart valve should ideally function like a natural heart valve.[11] The functioning of natural heart valves is characterized by many advantages:

  • Minimal regurgitation – This means that the amount of blood leaking backwards through the valve as it closes is small. Some degree of valvular regurgitation is inevitable and natural, up to around 5 ml per beat.[31] However, several heart valve pathologies (e.g. rheumatic endocarditis) may lead to clinically significant valvular regurgitation. A desirable characteristic of heart valve prostheses is that regurgitation is minimal over the full range of physiological heart function.
  • Minimal transvalvular pressure gradient – Whenever a fluid flows through a restriction, such as a valve, a pressure gradient arises over the restriction. This pressure gradient is a result of the increased resistance to flow through the restriction. Natural heart valves have a low transvalvular pressure gradient as they present little obstruction to the flow through themselves, normally less than 16 mmHg. A desirable characteristic of heart valve prostheses is that their transvalvular pressure gradient is as small as possible.
  • Non-thrombogenic – Natural heart valves are lined with an endothelium comparable with the endothelium lining the heart chambers, so they are not normally thrombogenic (i.e. they don't cause blood clots). Blood clots can be hazardous because they can lodge in, and block, downstream arteries (e.g. coronary arteries, leading to heart attack [myocardial infarction]; or cerebral arteries, leading to stroke). A desirable characteristic of artificial heart valves is that they are non- or minimally thrombogenic.
  • Self-repairing – Valve leaflets retain some capacity for repair thanks to regenerative cells (e.g. fibroblasts) in the connective tissue from which the leaflets are composed. As the human heart beats approximately 3.4×109 times during a typical human lifespan, this limited but nevertheless present repair capacity is critically important. No heart valve prostheses can currently self-repair, but tissue-engineered valves may eventually offer such capabilities.[26]

The performance of an artificial heart valve can be tested in vitro before clinical use by means of a pulse duplicator.[32]

Artificial heart valve repair

[edit]

Artificial heart valves are expected to last from 10 to 30 years.[16]

The most common problems with artificial heart valves are various forms of degeneration, including gross billowing of leaflets, ischemic mitral valve pathology, and minor chordal lengthening.[24] The repairing process of the artificial heart valve regurgitation and stenosis usually requires an open-heart surgery, and a repair or partial replacement of regurgitant valves is usually preferred.[24]

Researchers are investigating catheter-based surgery that allows repair of an artificial heart valve without large incisions.[33]

Researchers are investigating Interchangeable Prosthetic Heart Valve that allows redo and fast-track repair of an artificial heart valve. [34]

Additional images

[edit]

See also

[edit]

References

[edit]
  1. ^ Kostrzewa B, Rybak Z (2013). "[History, present and future of biomaterials used for artificial heart valves]". Polimery W Medycynie. 43 (3): 183–9. PMID 24377185.
  2. ^ Baumgartner H, Falk V, Bax JJ, De Bonis M, Hamm C, Holm PJ, et al. (September 2017). "2017 ESC/EACTS Guidelines for the management of valvular heart disease". European Heart Journal. 38 (36): 2739–2791. doi:10.1093/eurheartj/ehx391. PMID 28886619.
  3. ^ "Heart Valve Replacement: Which Type Is Best for You?". Health Essentials from Cleveland Clinic. 2018-06-14. Archived from the original on 2020-09-15. Retrieved 2020-08-04.
  4. ^ Muraru D, Anwar AM, Song JK (December 2016). "Heart valve disease: tricuspid valve disease". Oxford Medicine Online. doi:10.1093/med/9780198726012.003.0037.
  5. ^ "Surgical aortic heart valves | Edwards Lifesciences". www.edwards.com. Retrieved 2019-07-29.
  6. ^ Medtronic. "Heart Valve Therapies - Surgical Replacement". www.medtronic.com. Archived from the original on 2019-07-30. Retrieved 2019-07-29.
  7. ^ "Surgical Valve Solutions". Abbott. Retrieved 2024-07-22.
  8. ^ "On-X Heart Valves". CryoLife, Inc. Archived from the original on 2019-07-30. Retrieved 2019-07-29.
  9. ^ "Cardiac | LifeNet Health". www.lifenethealth.org. Archived from the original on 2019-04-19. Retrieved 2019-07-29.
  10. ^ a b Gott VL, Alejo DE, Cameron DE (December 2003). "Mechanical heart valves: 50 years of evolution". The Annals of Thoracic Surgery. 76 (6): S2230-9. doi:10.1016/j.athoracsur.2003.09.002. PMID 14667692.
  11. ^ a b c Pibarot P, Dumesnil JG (February 2009). "Prosthetic heart valves: selection of the optimal prosthesis and long-term management". Circulation. 119 (7): 1034–48. doi:10.1161/CIRCULATIONAHA.108.778886. PMID 19237674.
  12. ^ a b c Bloomfield P (June 2002). "Choice of heart valve prosthesis". Heart. 87 (6): 583–9. doi:10.1136/heart.87.6.583. PMC 1767148. PMID 12010950.
  13. ^ a b Matthews AM (1998). "The development of the Starr-Edwards heart valve". Texas Heart Institute Journal. 25 (4): 282–93. PMC 325574. PMID 9885105.
  14. ^ Goldsmith I, Turpie AG, Lip GY (November 2002). "Valvar heart disease and prosthetic heart valves". BMJ. 325 (7374): 1228–31. doi:10.1136/bmj.325.7374.1228. PMC 1124694. PMID 12446543.
  15. ^ Sun JC, Davidson MJ, Lamy A, Eikelboom JW (August 2009). "Antithrombotic management of patients with prosthetic heart valves: current evidence and future trends". Lancet. 374 (9689): 565–76. doi:10.1016/S0140-6736(09)60780-7. PMID 19683642. S2CID 43661491.
  16. ^ a b c d e f Tillquist MN, Maddox TM (February 2011). "Cardiac crossroads: deciding between mechanical or bioprosthetic heart valve replacement". Patient Preference and Adherence. 5: 91–9. doi:10.2147/PPA.S16420. PMC 3063655. PMID 21448466.
  17. ^ Johansen P (September 2004). "Mechanical heart valve cavitation". Expert Review of Medical Devices. 1 (1): 95–104. doi:10.1586/17434440.1.1.95. PMID 16293013. S2CID 27933945.
  18. ^ Namdari M, Eatemadi A (December 2016). "Nanofibrous bioengineered heart valve-Application in paediatric medicine". Biomedicine & Pharmacotherapy. 84: 1179–1188. doi:10.1016/j.biopha.2016.10.058. PMID 27780149.
  19. ^ Hickey GL, Grant SW, Bridgewater B, Kendall S, Bryan AJ, Kuo J, Dunning J (June 2015). "A comparison of outcomes between bovine pericardial and porcine valves in 38,040 patients in England and Wales over 10 years". European Journal of Cardio-Thoracic Surgery. 47 (6): 1067–74. doi:10.1093/ejcts/ezu307. PMID 25189704.
  20. ^ Harris C, Croce B, Cao C (July 2015). "Tissue and mechanical heart valves". Annals of Cardiothoracic Surgery. 4 (4): 399. doi:10.3978/j.issn.2225-319X.2015.07.01. PMC 4526499. PMID 26309855. Archived from the original on 2019-07-29. Retrieved 2019-07-29.
  21. ^ Johnston DR, Soltesz EG, Vakil N, Rajeswaran J, Roselli EE, Sabik JF, et al. (April 2015). "Long-term durability of bioprosthetic aortic valves: implications from 12,569 implants". The Annals of Thoracic Surgery. 99 (4): 1239–47. doi:10.1016/j.athoracsur.2014.10.070. PMC 5132179. PMID 25662439.
  22. ^ Flameng W, Hermans H, Verbeken E, Meuris B (January 2015). "A randomized assessment of an advanced tissue preservation technology in the juvenile sheep model". The Journal of Thoracic and Cardiovascular Surgery. 149 (1): 340–5. doi:10.1016/j.jtcvs.2014.09.062. PMID 25439467.
  23. ^ Bartuś K, Litwinowicz R, Kuśmierczyk M, Bilewska A, Bochenek M, Stąpór M, et al. (2017-12-19). "Primary safety and effectiveness feasibility study after surgical aortic valve replacement with a new generation bioprosthesis: one-year outcomes". Kardiologia Polska. 76 (3): 618–624. doi:10.5603/KP.a2017.0262. PMID 29297188. S2CID 4061454.
  24. ^ a b c d e f Hasan A, Saliba J, Pezeshgi Modarres H, Bakhaty A, Nasajpour A, Mofrad MR, Sanati-Nezhad A (October 2016). "Micro and nanotechnologies in heart valve tissue engineering". Biomaterials. 103: 278–292. doi:10.1016/j.biomaterials.2016.07.001. PMID 27414719.
  25. ^ a b c Stassen OM, Muylaert DE, Bouten CV, Hjortnaes J (September 2017). "Current Challenges in Translating Tissue-Engineered Heart Valves". Current Treatment Options in Cardiovascular Medicine. 19 (9): 71. doi:10.1007/s11936-017-0566-y. PMC 5545463. PMID 28782083.
  26. ^ a b c d Blum KM, Drews JD, Breuer CK (June 2018). "Tissue-Engineered Heart Valves: A Call for Mechanistic Studies". Tissue Engineering. Part B, Reviews. 24 (3): 240–253. doi:10.1089/ten.teb.2017.0425. PMC 5994154. PMID 29327671.
  27. ^ a b Theus AS, Tomov ML, Cetnar A, Lima B, Nish J, McCoy K, Mahmoudi M, Serpooshan V (2019-06-01). "Biomaterial approaches for cardiovascular tissue engineering". Emergent Materials. 2 (2): 193–207. doi:10.1007/s42247-019-00039-3. ISSN 2522-574X. S2CID 201165755.
  28. ^ Hay ED (2013-11-11). Cell Biology of Extracellular Matrix (Second ed.). Springer Science & Business Media. ISBN 978-1-4615-3770-0.
  29. ^ a b Nachlas AL, Li S, Davis ME (December 2017). "Developing a Clinically Relevant Tissue Engineered Heart Valve-A Review of Current Approaches". Advanced Healthcare Materials. 6 (24): 1700918. doi:10.1002/adhm.201700918. PMID 29171921. S2CID 2339263.
  30. ^ a b Schmidt JB, Tranquillo RT (2013). Tissue-Engineered Heart Valves. Boston, MA: Springer US. pp. 261–280. doi:10.1007/978-1-4614-6144-9_11. ISBN 978-1-4614-6143-2. {{cite book}}: |work= ignored (help)
  31. ^ Kasegawa H, Iwasaki K, Kusunose S, Tatusta R, Doi T, Yasuda H, Umezu M (January 2012). "Assessment of a novel stentless mitral valve using a pulsatile mitral valve simulator". The Journal of Heart Valve Disease. 21 (1): 71–5. PMID 22474745.
  32. ^ Mashari A, Knio Z, Jeganathan J, Montealegre-Gallegos M, Yeh L, Amador Y; et al. (2016). "Hemodynamic Testing of Patient-Specific Mitral Valves Using a Pulse Duplicator: A Clinical Application of Three-Dimensional Printing". Journal of Cardiothoracic and Vascular Anesthesia. 30 (5): 1278–85. doi:10.1053/j.jvca.2016.01.013. PMID 27179613.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  33. ^ Bezuidenhout D, Williams DF, Zilla P (January 2015). "Polymeric heart valves for surgical implantation, catheter-based technologies and heart assist devices". Biomaterials. 36: 6–25. doi:10.1016/j.biomaterials.2014.09.013. PMID 25443788.
  34. ^ Roberto de Menezes Lyra | title = Eight Types of Interchangeable Prosthetic Heart Valve: A Mini Review. | journal = https://globaljournals.org Archived 2023-03-22 at the Wayback Machine | https://globaljournals.org/GJMR_Volume22/2-Eight-Types-of-Interchangeable-Prosthetic-Heart-Valve.pdf Archived 2023-03-22 at the Wayback Machine

Further reading

[edit]
  • Bendet I, Morozov SM, Skumin VA [in French] (June 1980). "[Psychological aspects of the rehabilitation of patients after the surgical treatment of heart defects]" Психологические аспекты реабилитации больных после хирургического лечения пороков сердца [Psychological aspects of the rehabilitation of patients after the surgical treatment of heart defects]. Kardiologiia (in Russian). 20 (6): 45–51. PMID 7392405.
  • Skumin VA [in French] (September 1979). "[Nurse's role in medico-psychological rehabilitation of patients with artificial heart valves]". Meditsinskaia Sestra. 38 (9): 44–5. PMID 259874.
  • Skumin VA [in French] (1982). "[Nonpsychotic mental disorders in patients with acquired heart defects before and after surgery (review)]". Zhurnal Nevropatologii I Psikhiatrii imeni S.S. Korsakova. 82 (11): 130–5. PMID 6758444.
  • Klepetko W, Moritz A, Mlczoch J, Schurawitzki H, Domanig E, Wolner E (January 1989). "Leaflet fracture in Edwards-Duromedics bileaflet valves". The Journal of Thoracic and Cardiovascular Surgery. 97 (1): 90–4. doi:10.1016/S0022-5223(19)35130-X. PMID 2911200.
  • Podesser BK, Khuenl-Brady G, Eigenbauer E, Roedler S, Schmiedberger A, Wolner E, Moritz A (May 1998). "Long-term results of heart valve replacement with the Edwards Duromedics bileaflet prosthesis: a prospective ten-year clinical follow-up". The Journal of Thoracic and Cardiovascular Surgery. 115 (5): 1121–9. doi:10.1016/s0022-5223(98)70412-x. PMID 9605082.
  • Knapp RJ, Daily JW, Hammitt FG (1970). Cavitation. New York: McGraw-Hill Int. Book Co.
  • Lim WL, Chew YT, Low HT, Foo WL (September 2003). "Cavitation phenomena in mechanical heart valves: the role of squeeze flow velocity and contact area on cavitation initiation between two impinging rods". Journal of Biomechanics. 36 (9): 1269–80. doi:10.1016/s0021-9290(03)00161-1. PMID 12893035.
  • Bluestein D, Einav S, Hwang NH (November 1994). "A squeeze flow phenomenon at the closing of a bileaflet mechanical heart valve prosthesis". Journal of Biomechanics. 27 (11): 1369–78. doi:10.1016/0021-9290(94)90046-9. PMID 7798287.
  • Graf T, Fischer H, Reul H, Rau G (March 1991). "Cavitation potential of mechanical heart valve prostheses". The International Journal of Artificial Organs. 14 (3): 169–74. doi:10.1177/039139889101400309. PMID 2045192. S2CID 23086590.
  • Kafesjian R, Wieting DW, Ely J, Chahine GL, Frederick GS, Watson RE (1990). "Characterization of Cavitation Potential of Pyrolitic Carbon". In Bodnar E (ed.). Surgery for Heart Valve Disease: Proceedings of the 1989 Symposium. ICR. pp. 509–16. ISBN 978-1-872743-00-4.
  • Chahine GL (March 1996). "Scaling of mechanical heart valves for cavitation inception: observation and acoustic detection". The Journal of Heart Valve Disease. 5 (2): 207–14, discussion 214–5. PMID 8665016.
  • Zapanta CM, Stinebring DR, Sneckenberger DS, Deutsch S, Geselowitz DB, Tarbell JM, et al. (1996). "In vivo observation of cavitation on prosthetic heart valves". ASAIO Journal. 42 (5): M550-5. doi:10.1097/00002480-199609000-00047. PMID 8944940.
  • Richard G, Beavan A, Strzepa P (April 1994). "Cavitation threshold ranking and erosion characteristics of bileaflet heart valve prostheses". The Journal of Heart Valve Disease. 3 (Suppl 1): S94-101. PMID 8061875.
[edit]